Please check the standard weighing param-
eters:
Menu a Weighing a Overview
Adapt filter:
Stable conditions
Application filter:
Final readout
Stability:
Maximum speed
Stability delay:
Very short delay
After stability
Zero/tare function:
Back
Wizard
Please check the standard weighing param-
eters:
Menu a Weighing a Overview
Zero/tare function:
After stability
Automatic zeroing:
Off
Basic unit:
Milligrams
Display accuracy:
Show all decimal places
Zero/tare at power on:
On
Back
Wizard
Device parameters
Configure balance parameters
Leveller configuration
Please check the leveling parameters:
Menu a Leveling a Overview
Display
Leveling notice:
Back
Adapt filter
: The effects of unfavorable weighing conditions such as drafts or
vibration can be filtered out by adapting the balance filter (and thus its measuring
time) to suit ambient conditions. Options:
/
conditions
Unstable conditions
Application filter
Options: *
Final readout
Stability
: The stability indicator lights up when the weighing results are
constant within a set range. Options:
/*
accuracy
Medium accuracy
Save
Stability delay
such as turbulence in the weighing chamber. Options:
delay/Medium delay/Long delay
Zero/tare function:
options:
Without stability
Automatic zeroing
fraction of scale intervals per second starting from the display zero point are
automatically tared. Options:
Basic unit
: Weight unit to be used for this task; options:
unit
/*
milligrams
Display accuracy
Options: *
All dig. on
Save
/
change
Last decimal place of the 1st Division
Zero/Tare at Power On
tared when switched on.
t
Touch
to save your settings.
Save
4.5.1.2 Setting up Q-Level Leveling Function
Menu
You can configure the leveling function here.
Leveling notice
Switched off/Switching off, no indication
Display/Status Display
appears and flashes.
Alert/Alarm message, Leveling mandatory
the
Level me!
appears after 5 minutes. If this occurs, work can only be resumed after leveling has
been performed.
Save
Touch
to save your settings.
Save
Very stable conditions
/
Very unstable conditions
: Equalizes the load deviations on the display.
/
Filling mode
/
Low filtering
Maximum accuracy
/
/
Fast
Very Fast
: This setting lets you compensate for slowly abating disturbances,
Conditions for performing the tare function;
/*After stability
: If this option is enabled, changes of a fixed, configured
Switching on
: A low display accuracy gives a faster display.
/
Reduced by one decimal place in case of load
:
The
pipette calibration system
Options: *
Switching on
: As soon as leveling is required the
tab appears and flashes. If this notice is ignored, an alarm message
Cubis
System Settings (Menu)
/*
Stable
/
Without filtering
/
Very high
/
Maximum speed
Very short delay/Short
/During stability
/*
Switching off
Any available weight
/
Last decimal place off
is automatically
/
switching off
: Leveling prompt is disabled.
Level me!
: As soon as leveling is required
®
MPS Quick Reference Guide