Configuring Calibration/Adjustment; Defining Calibration And Adjustment Functions; Defining External Calibration Weights - Sartorius cubis MPS Quick Reference Manual

Electronic pipette calibration system
Table of Contents

Advertisement

Please select the menu item:
Menu
Level balance
Calibration/adjustment data
Configure timer controlled actions
Display device information
Device parameters
Back
Language
Please configure the calibration/adjust-
ment functions:
Menu a Cal./Adj.
Define calibration/adjustment functions
Define external calibration weights
Define external linearization weights
Set fully autom. adjustment function isoCAL
Back
Calibration/adjustment data
Define calibration/adjustment functions
Calibration/adjustment data
Define external calibration weights
4.2

Configuring Calibration/Adjustment

t
In the operating mode, touch
y
The selection of the system settings is displayed.
t
Touch
Calibration/adjustment data
y
The selection of the settings options is displayed (the settings options will be
described in the following chapters).
− Define calibration/adjustment functions
− Define external calibration weights
− Define external linearization weights
− Set fully autom. adjustment function isoCAL
t
Select the section in which you want to make settings.
t
Configure the desired settings for all sub-menus.
t
To save the settings, touch
Next
In the following text, factory settings are identified with an asterisk (*).
4.2.1

Defining Calibration and Adjustment Functions

t
In the menu, touch
the settings options.
t
In the selection section of the settings options, touch
adjustment functions
Cal./Adj. key
Select fixed 1-key function
Cal./Adj. Selection menu
are listed when the Cal./Adj. key is touched.
Cal./Adj. Sequence
then adjust manually
Cal./Adj. Unit
4.2.2

Defining External Calibration Weights

t
In the menu, touch
the settings options.
t
In the selection of the settings options, touch
weights
:
No. of ext. cal. weights
Ext. cal. weight n
Weight value (in grams), ID, certificate, validation
Menu
.
.
Save
.
Calibration/adjustment data
:
: Functions of the key: Block key/*Configure key selection menu/
: Select which calibration/adjustment functions
: Options: *Calibrate, then adjust automatically/calibrate,
: Options: *Gram/Kilogram/Defined via data record
Calibration/adjustment data
: Enter the number
: Enter the data for each calibration weight:
®
Cubis
System Settings (Menu)
to access the selection of
Define calibration/
to access the selection of
Define external calibration
MPS Quick Reference Guide
17