Sartorius cubis MPS Quick Reference Manual page 20

Electronic pipette calibration system
Table of Contents

Advertisement

System Settings (Menu)
Please enter the execution date and time:
...
a New a Planner a Wizard a Execution date/time
Date:
Time:
Back
Overview
Info
Please enter the repetition time period:
Actions a New a Planner a Wizard a Repetition time
Years:
Months:
Days:
Hours:
Minutes:
Back
Overview
Info
Please select action execution:
Actions a New a Planner a Wizard a Execution
Enable action
Disable action
Back
Overview
Check parameters for timer controlled
actions:
Actions a New a Planner a Overview
Action execution:
Message and function
Function:
Internal calibration/adjustment
Message text:
Timer controlled adjustment
Execution date/time:
02/23/2016; 12:00:00
Repetition time:
0 ; 0 ; 1 ; 0 ; 0
Back
Wizard
Please select action:
Actions
Action 1
02/23/2016 12:00:00; Y:0 M:0 D:1 S:0 M:0
®
20
Cubis
MPS Quick Reference Guide
t
Enter the date and time at which the action should be executed for the first time.
02/23/2016
12:00:00
Next
t
Define the time interval at which you want the action to be repeated (e.g., 1 day
for daily occurrence).
t
Touch
Next
Next
t
Select
Enable action
y
The programmed action will start at the specified time. You can at any time
enable or disable the window "Enable/Disable Action" in order to disable the
programmed action or to enable it once again with a new start time.
Next
y
The overview of all settings for the action is displayed.
t
Check and correct all settings as required.
t
Touch
Next
t
Enter a name and a description for the action.
t
Touch
Save
t
Touch
Back
Next
y
In the operating mode, a tab "Timer" is displayed. If you touch the tab, you can
Timer
access a list of the actions.
y
The programmed action is displayed in the list of actions. You can edit them
again at any time here.
activated
.
.
.
.
.

Advertisement

Table of Contents
loading