Using The X-Ray Tray - Dräger Isolette 8000 plus Instructions For Use Manual

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

1 Pre-warm the incubator.
2 Rotate the locking knobs, and open the front access panel.
3 Place the infant in the center of the bed.
4 Close the access panel, and ensure that the locking knobs are fully engaged.
8.11

Using the X-ray tray

WARNING
Risk of death or serious injury
Unattended infants could roll out of the incubator.
► Do not leave the infant unattended when the access panel is open.
WARNING
Risk of death or serious injury
When an X-ray is taken through the hood, the hood could show up on the X-ray as
a translucent shadow and could result in incorrect diagnosis.
► Be aware of this possibility when taking X-rays through the hood.
1 Remove any accessories that may interfere with movement of the front access
panel.
2 Rotate the locking knobs, and open the front access panel.
3 Slide out the X-ray tray (A) from under the bed.
4 Place the X-ray cassette in the center of the X-ray tray.
5 Use the numbered scales on the X-ray tray and on the hood to position the
patient and the X-ray cassette.
6 Push the X-ray tray back in under the bed.
7 Place the infant at the center of the bed.
8 Close the access panel, and ensure that the locking knobs are fully engaged.
9 When the X-ray is completed, repeat steps 1 and 2.
10 Remove the X-ray cassette from the tray, and return the tray.
11 Close the access panel, and ensure that the locking knobs are fully engaged.
12 Replace any accessories that were previously removed.
Instructions for use Isolette 8000 plus SW 5.n
A
Operation
105