Trigger Coupling - Rigol MSO2000A Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

Chapter 5 To Trigger the Oscilloscope

Trigger Coupling

Trigger coupling decides which kind of components will be transmitted to the trigger
module. Please distinguish it with "Channel Coupling".
DC: allow DC and AC components into the trigger path.
AC: block all the DC components and attenuate signals lower than 75 Hz.
LF Reject: block the DC components and reject the low frequency components
(lower than 75 kHz).
HF Reject: reject the high frequency components (higher than 75 kHz).
Press MENU  Setting  Coupling in the trigger control area (TRIGGER) at the
front panel to select the desired coupling type (the default is DC).
Note:
CH1 or CH2.
MSO2000A/DS2000A User's Guide
nd Quality Products Online at:
Trigger coupling is only valid in edge trigger and when the trigger source is
GlobalTestSupply
www.
.com