Setup/Hold Trigger

In setup/hold trigger, you need to set the data signal line and clock signal line. The
setup time starts when the data signal passes the trigger level/threshold level and
ends at the coming of the specified clock edge; the hold time starts at the coming of
the specified clock edge and ends when the data signal passes the trigger
level/threshold level again (as shown in the figure below). During the setup or hold,
the data signal must remain a valid pattern. The oscilloscope triggers when the setup
time or hold time is lower than the preset time.
Data Source
Clock Source
Trigger Type:
Press Type to open the trigger type list. Rotate
"Setup/Hold". At this point, the trigger setting information as shown in the figure
below is displayed at the upper right corner of the screen. For example,
the trigger level is -1.80 V.
Source Selection:
Press DataSrc and ClkSrc to set the signal sources of the data line and clock line
respectively. They can be set to CH1, CH2 or any channel of D0-D15. For the details,
please refer to the introduction in "Trigger Source". The current trigger source is
displayed at the upper right corner of the screen.
Note:
Select channel with signal input as trigger source to obtain stable trigger.
5-36
nd Quality Products Online at:
T2
T1
The data type is H
The edge type is
Figure 5-12 Setup/Hold Trigger
. The trigger type is setup/hold trigger; the trigger source is CH1;
GlobalTestSupply
www.
Chapter 5 To Trigger the Oscilloscope
T1 is the setup time
T2
is the hold time
and press down
MSO2000A/DS2000A User's Guide