Vertical Scale - Rigol MSO2000A Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

RIGOL

Vertical Scale

Vertical scale refers to the voltage value per grid in the vertical direction on the
screen and is usually expressed as V/div.
Press CH1  Volts/Div to select the desired adjustment mode of the vertical scale,
or press VERTICAL
Coarse adjustment: set the vertical scale in 1-2-5 step namely 500 μV/div, 1
mV/div, 2 mV/div, 5 mV/div, 10 mV/div...10 V/div (take counterclockwise as an
example).
Fine adjustment: further adjust the vertical scale at a relatively smaller step to
improve the vertical resolution. If the amplitude of the input waveform is a little
bit greater than the full scale under the current scale and the amplitude would
be a little bit lower if the next scale is used, fine adjustment can be used to
improve the amplitude of waveform display to view signal details.
After the adjustment mode is selected, rotate VERTICAL
vertical scale (clockwise to reduce the scale and counterclockwise to increase). The
scale information in the channel status label (as shown in the figure below) at the
bottom of the screen will change accordingly during the adjustment. The adjustable
range of the vertical scale is related to the probe ratio and input impedance currently
set. By default, the probe ratio is 1X, the input impedance is 1 MΩ and the adjustable
range of the vertical scale is from 500 μV/div to 10 V/div.
When adjusting the vertical scale, the waveform will be expanded or compressed
around the center of the screen or the ground level of the signal according to the
change of the vertical scale.
2-6
nd Quality Products Online at:
SCALE to switch the adjustment mode quickly.
GlobalTestSupply
www.
Chapter 2 To Set the Vertical System
SCALE to adjust the
MSO2000A/DS2000A User's Guide
.com
sales@GlobalTestSupply.co

Advertisement

Table of Contents
loading

Table of Contents