Chapter 5 To Trigger the Oscilloscope

Trigger Coupling

Trigger coupling decides which kind of components will be transmitted to the trigger
module. Please distinguish it with "Channel Coupling".
DC: allow DC and AC components into the trigger path.
AC: block all the DC components and attenuate signals lower than 75 Hz.
LF Reject: block the DC components and reject the low frequency components
(lower than 75 kHz).
HF Reject: reject the high frequency components (higher than 75 kHz).
Press MENU  Setting  Coupling in the trigger control area (TRIGGER) on the
front panel to select the desired coupling type (the default is DC).
Note:
Trigger coupling is only valid in edge trigger and when the trigger source is
CH1 or CH2.