Trigger Output Connector - Rigol MSO2000A Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO2000A Series:
Table of Contents

Advertisement

Chapter 5 To Trigger the Oscilloscope
RIGOL

Trigger Output Connector

The trigger output connector ([Trigger Out]) at the rear panel of the oscilloscope
can output trigger signals determined by the current setting.
Trigger Output Connector
The oscilloscope can output a signal that reflects the current capture rate of the
oscilloscope from the [Trigger Out] connector each time a trigger is generated.
Connect the signal to a waveform display device and measure the signal frequency;
the measurement result is the same with the current capture rate.
Note:
If Utility  AuxOutput is set to "PassFail" or Utility  Pass/Fail 
AuxOutput is set to "ON", during the pass/fail test, the oscilloscope will output a
negative pulse through the [Trigger Out] connector each time a failure is detected;
the oscilloscope will output low level continuously through the [Trigger Out]
connector when no failure is detected.
MSO2000A/DS2000A User's Guide
5-53