Chapter 10 Pass/Fail Test
outputs a pulse from the [Trigger Out] connector at the rear panel when a
failed waveform is detected.
You can also press Utility  AuxOutput and select "PassFail" to set this
output.

To Save the Test Mask

Users can save the current test mask in "*.pf" format to the internal Flash memory or
external USB storage device. The internal memory can store at most 10 test mask
files (LocalPF.pf).
Press Save to enter the file store interface. Please refer to the relative descriptions in
"Store and Recall" to save the test mask file to the internal or external memory.

To Load the Test Mask

Users can also load the test mask files (*.pf) stored in the internal Flash memory or
external USB storage device to the instrument.
Press Load to enter the file recall interface. Please refer to the relative descriptions
in "Store and Recall" to load the test masks to the instrument.
MSO2000A/DS2000A User's Guide