Chapter 6 MATH and Measurements
Note:
automatically and cannot be set.
When the display mode is "Y" or "X-Y", press Vertical Unit to select "Source
Unit" or "%".
Note:
LA, the Vertical Unit menu will be hidden automatically and cannot be set.
When the measurement source is "LA"
When the measurement source is "LA", DisplayMode is set to "X" automatically
and cannot be changed; you can only set the X axis unit. Press Time Unit
softkey to select "s", "Hz", "°" or "%".
MSO2000A/DS2000A User's Guide
nd Quality Products Online at:
°: in the measurement results, A->X, B->X and △X are expressed in degree.
At this point, A->X, B->X and △X will change to "0°", "360°" and "360°"
respectively when you press SetCursor, no matter where cursor A and
cursor B are currently located. At the same time, two cursor lines
(unmovable) appear on the screen as the reference positions.
%: in the measurement results, A->X, B->X and △X are expressed in
At this point, A->X, B->X and △X will change to "0%", "100%" and "100%"
percentage.
respectively when you press SetCursor, no matter where cursor A and B
are currently located. At the same time, two cursor lines (unmovable)
appear on the screen as the reference positions.
When the display mode is set to "Y", the Time Unit menu will be hidden
units of A->Y, B->Y and △Y will be automatically set to the unit of the
Source Unit: when this unit is selected, in the measurement results, the
current source.
△X are expressed in percentage. At this point, A->Y, B->Y and △X will
%: when this unit is selected, in the measurement results, A->Y, B->Y and
change to "0%", "100%" and "100%" respectively when you press
SetCursor, no matter where cursor A and B are currently located. At the
same time, two cursor lines (unmovable) appear on the screen as the
reference positions.
When the display mode is set to "X" or the measurement source is set to
s: in the measurement results, CurA, CurB and △X denote time.
Hz: in the measurement results, CurA, CurB and △X denote frequency.
°: in the measurement results, CurA, CurB and △X are expressed in degree.
For the description and setting method of phase reference position, please
refer to the related introduction above.
GlobalTestSupply
www.
.com
sales@GlobalTestSupply.co
RIGOL
6-33