To Save The Test Mask Range; To Load The Test Mask Range - Rigol DS4024E User Manual

Ds4000e series; digital oscilloscope
Hide thumbs Also See for DS4024E:
Table of Contents

Advertisement

Chapter 9 Pass/Fail Test

To Save the Test Mask Range

When the pass/fail test function is enabled, you can save the current test mask range
to the internal memory or external USB storage device (only when a USB storage
device is detected) in "*.pf" format. The internal memory can store at most 10 test
mask files (LocalPF0.pf to LocalPF9.pf).
Note: For the test mask range, please refer to "To Set the Test Mask".
1.
Press Utility  Pass/Fail  Enable and select "ON" to enable the pass/fail
test function.
Press Save to enter the file store interface. Please refer to the relative
2.
descriptions in "Store and Recall" to save the test mask file to the internal or
external memory.

To Load the Test Mask Range

When the pass/fail test function is enabled, you can load the test mask file stored in
the internal memory or external USB storage device (only when a USB storage device
is detected) and apply it to the current pass/fail test function.
Press Utility  Pass/Fail  Enable and select "ON" to enable the pass/fail
1.
test function.
2.
Press Load to enter the file recall interface. Please refer to the relative
descriptions in "Store and Recall" to load the specified test mask file (*.pf)
and apply it to the current pass/fail test function.
DS4000E User's Guide
RIGOL
9-9