7.6.1 Setup

Reactive Values Calculation:
culation.
Nominal Frequency:
coefficients used for calculating power and energy.
Current Sensor:
or ADA Adapter (used to accept probes with other ratios or a direct 1 Amp or
5 Amp input)
Connection Type:
or 3-Phase 5 wire.
Set Instrument's Clock:
of the computer or a desired setting.
Voltage Transformer Ratio:
where measurements are on the secondary side of a transformer and the pri-
mary value needs to be displayed.
76
Figure 7-7
With or without harmonics. Applies to VAR cal-
50 or 60Hz. This parameter determines the correction
MN93, MN193, SR193, MR193, A193 AmpFlex
Single-Phase, Two-Phase (Split Phase), 3-Phase 3 wire
Resets the time and date on the PowerPad
Sets the scale for voltage measurement in cases
Power Quality Analyzer Model 8335