3.1 Reference Conditions

Parameter
Ambient temperature
Humidity
Atmospheric pressure
Phase voltage
Clamp current circuit input voltage
AmpFlex
current circuit input voltage
®
Frequency of electricity network
V/I phase shift
Harmonics
Voltage unbalance
NOTE: The symbol "U" will be used throughout this manual and in the
instrument to refer to phase-to-phase voltage measurement.
The symbol "V" will be used for phase-to-neutral voltage measurement.
*Specifications are subject to change without notice.

3.2 Electrical Specifications

Sampling Frequency (256 samples per cycle):
12.8kHz samples/sec per channel @ 50Hz
15.36kHz samples/sec per channel @ 60Hz
Waveforms: Displays voltages and currents
Screen captures: 50 max
Transients: Detection and recording of transients (up to 210)
Inrush Current: Detection and recording of Inrush current (1 max)
Memory: 2GB with date and time programming
Alarm Function: 10,000 alarms max
16
CHAPTER 3

SPECIFICATIONS

Reference Conditions
73°F ± 5°F (23°C ± 3°C)
45% to 75%
25.4" Hg to 31.3" Hg (860 to 1060 hPa)
50Vrms to 1000Vrms without DC (<0.5%)
30mVrms to 1Vrms without DC (<0.5%)
11.73mVrms to 117.3mVrms without DC (<0.5%)
50 and 60Hz ± 0.1Hz
0° active power / 90° reactive power
<0.1%
<10%