3.3 Mechanical Specifications
Dimensions: 7.9 x 9.8 x 2.6" (200 x 250 x 67mm)
Weight: 4.3 lb (1950g) with batteries
Shock and Vibration: per EN 61010-1
Tightness: IP 50 per EN 60529 (electrical IP2X for terminals)
3.4 Environmental Specifications
Altitude: Operating: 0 to 2000 meters (6560 ft)
Non-Operating: 0 to 10,000 meters (32800 ft)
Temperature and % RH:
95
90
75
45
10
0
3.5 Safety Specifications
Electrical Safety
EN 61010-1; 600V CAT IV* (probe dependent); Pollution Degree 2
Indoor Use
*Warning: The assigned voltage and measuring category of the "device + current
probe" may differ from the characteristics of the device alone.
•
The use of the AmpFlex
probe" at 600V CAT IV or 1000V CAT III.
•
The use of the MR193, MN93 and MN193 probes rate the "device + cur-
rent probe" at 300V CAT IV or 600V CAT III.
•
The use of the 5A adapter rates the "device + current probe" at
CAT IV or 300V CAT III.
24
-4
32
68
Temperature in °F
, MiniFlex™ and SR probes rate the "device +
®
79
95
122
Power Quality Analyzer Model 8335
Reference Range
Operating Range
Storage Range
150V