AEMC 8335 User Manual page 21

Powerpad 3-phase power quality analyzer
Hide thumbs Also See for 8335:
Table of Contents

Advertisement

Function
Phase Voltage TRMS
Half-period
Composite Voltage TRMS
Half-period
Peak Factor
Excluding
AmpFlex
& MiniFlex
Active Power
AmpFlex
& MiniFlex
Excluding
Reactive
AmpFlex
Powers
& MiniFlex
Inductive
&
AmpFlex
Capacitive
& MiniFlex
Apparent Power
Power Factor
(1) In 1000V
, Cat. III, providing that voltage between each of the terminals and ground does not exceed 1000V
RMS
(2) Two phase (opposite phases) – same note as (1).
(3) Caution: The absolute value of offset must not exceed 95% of peak amplitude.
In other words, s(t) = S × sin(ωt) + O, we have |O| ≤ 0.95 × S (with positive S).
MAX and MIN values in wave shape mode and the values V
current modes are half-period values.
NOTE: The uncertainties given for power and energy measurements are maximum for |Cos φ| = 1 or |Sin φ| = 1 and are
typical for other phase displacement values.
Power Quality Analyzer Model 8335
Measurement Range
Minimum
10V
(3)
10V
(3)
1
4
0Ω
®
®
0Ω
0VAR
®
®
0VAR
0VA
-1
Resolution
Maximum
0.1V
V < 1000V
1000V
(1)
1V
V ≥ 1000V
0.1V
U < 1000V
2000V
(2)
1V
U ≥ 1000V
3.99
0.01
9.99
0.01
1V
9999kW
U ≥ 1000V
9999kW
4 digits
9999kVAR
4 digits
9999kVAR
4 digits
9999kVA
4 digits
1
0.001
A
(excluding neutral channels) in Alarm and Inrush
RMS and
RMS
Accuracy
±(0.8% + 1V)
±(0.8% + 1V)
±(1% + 2cts)
±(5% + 2cts)
±(1%)
Cos φ ≥ 0.8
±(1.5% + 10cts)
0.2 ≤ Cos φ < 0.8
±(1%)
Cos φ ≥ 0.8
±(1.5% + 10cts)
0.5 ≤ Cos φ < 0.8
±(1%)
Sin φ ≥ 0.5
±(1.5% + 10cts)
0.2 ≤ Sin φ < 0.5
±(1.5%)
Sin φ ≥ 0.5
±(2.5% + 20cts)
0.2 ≤ Sin φ < 0.5
±(1%)
±(1.5%)
Cos φ ≥ 0.5
±(1.5% + 10cts)
0.2 ≤ Cos φ < 0.5
RMS.
19

Advertisement

Table of Contents
loading

Table of Contents