6-6
Recording in 6 Lead Mode: Recording
The locations of all relevant device settings are shown below:
Operating mode
Muscle filter
2
Report function
Failure message
1. if enabled
2. or REC OFF when the recording function is disabled
3. if any
If you change the writer speed, lead group or any filter settings
during a recording, the unit will briefly stop.
With
you advance to the next group of 6 leads of the selected
lead
report sequence.
When the anti-drift system is enabled, there will be a short delay of
2.2 seconds before the recording starts.
The heart rate limit is automatically calculated from the date of birth
(WHO 100% = 220 - age). When the date of birth is not entered, the unit
will set the limit at 180 bpm. This value can be changed with
(in steps of 5 bpm).
MAC 1200
2012250-022
1
1
AC line filter
Anti-drift system
Gain
Three leads
display here
3
1
"Patient Name"
Report sequence
Heart rate
and
Revision B