Muscle Filter Frequency

Trend Recording

Arrhythmia Data

Episodes

Revision B
System Setup: Arrhythmia Mode
Cut-off frequency of the muscle filter: [40 Hz], 20 Hz.
The frequency range is indicated in the lower margin of the recording
strip.
0.08 – 40 Hz (40-Hz muscle filter enabled)
0.08 – 20 Hz (20-Hz muscle filter enabled)
0.08 – 100 Hz (muscle filter off).
The slow trend recording of 5 mm/s automatically begins at program
start ([no]/yes).
The recorder will document arrhythmias in the following situations (all
[unequal], no).
All – the recorder documents all arrhythmias that occur.
Unequal – the recorder documents every arrhythmia that occurs that
is different from the preceding event.
No – arrhythmias are not documented.
Final report includes episode report, with episodes listed by one of the
following criteria ([chron.], prio., ventr., no.).
In chronological order
According to priorities (see Table 3,
Ventricular beats only
No episode report
MAC 1200
2012250-022
"Arrhythmia Codes,"
on page 7-9)