Force Technician Name - GE Medical Systems MAC 1200 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Memory Used
Untransmitted ECGs
419270-471680 bytes
471680+ bytes
1. The system will allow users to acquire (print) ECGs beyond the 40th ECG, but these subsequent ECGs will not be stored. The
Memory is full message will display at each ECG acquisition after 40.

Force Technician Name

Revision B
System Setup: High Security
On-screen Message
32-39
Warning: approaching ECG storage limit,
Transmit and delete ECGs to free memory
To close this message, Press key: <--'
40
Memory is full
When enabled, the Force Technician Name high security option requires
users to enter a technician name whenever an ECG test is initiated. If
this option is enabled, a test will not start if the technician name is not
entered.
The default value for Force Technician Name is Yes.
MAC 1200
2012250-022
Message Displays When...
When ECGs 32-39 are acquired
Every power up
When ECG 40 and all
subsequent ECGs are acquired
Every power up
1
9-19