AT Command for Establishing a Communication Link
AT Command for Termination of the Communication
Revision B
Recording in 12 Lead Mode: ECG Transmission
The following is an example of a dial string for a modem connected to a
branch (PBX system) and dialing a modem via the public telephone
network, using the touch tone mode.
Symbol
AT
Prefix that precedes every command line
DT
Touch tone dial mode
xxx
After DT, enter the characters for access to the public telephone network
(e.g. 0)
W
W, placed after a number, tells the modem in a PBX system to wait for the
dial tone of an outside telephone line
dial string:
ATDTOW
The communication is terminated in two steps.
First, the MAC 1200 sends an escape command to return from the on-
line state to the command state. Then the hangup command follows.
Symbol
+++
Escape command
AT
Prefix that precedes every command line
H
Hangup command
hangup string:
+++ATH