Location; Date/Time; Lead Fail Beep; High Hr Beep - GE Medical Systems MAC 1200 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Location

*

Date/Time

Lead Fail Beep

High HR Beep

Lead Labels

Pace Enhancement

Baseline Roll Filter

Revision B
System Setup: System Setup
This is a 3-digit ID number of the location to which the ECG recorder is
assigned. The default is 1, the range is 1 to 600.
Enter date and time (enter 4 digits for the year).
Audio signal to indicate electrode problems (yes, [no]).
Audio signal to indicate that the heart rate exceeds the limit value: yes,
[no] (only in 6 Lead and Arrhythmia modes). The limit value (220 - age)
can be changed manually.
1. The value selected here is the default value appearing in the patient
data
2. with CSI protocol only
[AAMI] codes: RA, LA, RL, LL, V1 to V6 or
IEC codes: R, L, F, N, C1 to C6
Pacer pulses are displayed enhanced and with a given amplitude ([yes],
no).
When programming pacemakers, we recommend to disable the pace
enhancement.
Select the lower cut-off frequency: = 0.04, [0.08], 0.16 Hz.
MAC 1200
2012250-022
9-21