Wiring The Power Supply - Emerson Micro Motion 5700 Installation Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Installation Manual
MMI-20039473
5

Wiring the power supply

You can install a user-supplied switch in the power supply line.
Important
For compliance with the Low Voltage Directive 2014/35/EU (European installations), verify
that the transmitter is located in close proximity to a switch.
Procedure
1. Remove the wiring access cover.
2. Open the Power warning flap to locate the power terminals.
3. Connect the power supply wires.
4. Tighten the two screws holding the power connector in place.
5. Ground the power supply using the equipment ground, also under the Power
Installation Manual
Figure 5-1: Location of power supply wiring terminals and equipment ground
A. Power supply wiring terminals (+ and -)
B. Conduit connection for power supply wiring
C. Equipment ground
For DC power: connect to terminals + and –.
For AC power: connect to terminals L/L1 (line) and N/L2 (neutral).
NOTICE
Do not use the conduit specified for power supply wires for input/output wiring (see
Figure
5-1) to avoid pinched wires when the cover is closed.
warning flap.
Wiring the power supply
October 2022
A
B
C
35