7.1
Clean the instrument externally
Maintenance interval
Clean the instrument externally when required. Do not allow spilled liquid to dry on the
instrument.
Required material
The following materials are required:
•
•
Instruction
Follow the steps below to clean the instrument externally.
Step
1
2
3
4
ÄKTA go Operating Instructions 29360951 AC
CAUTION
Disconnect the power cord form the power outlet before cleaning
the instrument.
Cloth
Mild cleaning agent or 20% ethanol
Action
Turn off the instrument and disconnect the power cord from the wall socket.
Wipe the surface with a damp cloth. Wipe off stains using a mild cleaning
agent or 20% ethanol. Wipe off any excess.
Remove the top and bottom tray for cleaning. Put them back once they are
cleaned.
Let the instrument dry completely before using it.
7 Maintenance
7.1 Clean the instrument externally
91