3.6
Packing flow rates
Table 3.4: Packing flow rates of AxiChrom columns.
Column
AxiChrom 50
AxiChrom 70
AxiChrom 100
The flow rate needed into the hydraulic chamber to move the adapter downwards at a
given linear velocity is lower than the flow rate needed to achieve a linear velocity through
the bed. This is due to the adapter rod volume in the hydraulic chamber.
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
Packing flow rate at 30 cm/h
8.8 ml/min
17.7 ml/min
35.8 ml/min
3 Specifications and characteristics
3.6 Packing flow rates
Packing flow rate at 60 cm/h
17.7 ml/min
35.4 ml/min
71.6 ml/min