Table 3.1 Error Messages (Continued)
Message
UNABLE TO COPY - MAXIMUM
NUMBER OF AMORPHOUS REGIONS
Unable to create the histogram queue
Unable to create the regions display
Unable to find/open the protocol file
Unable to find the Listmode file
Unable to find the Listmode file,
Carousel aborted on tube # [tube
number]
Unable to find the Listmode file,
Current operation aborted
Unable to find the panel info
Unable to get the available disk space
Unable to get the specified Cytometer
state
Unable to initialize histogram memory
Unable to initialize protocol memory
Unable to load current panel
PN 4237296CA
Cause
When copying an amorphous region,
the system is unable to complete the
task since the protocol already has the
maximum number of eight amorphous
regions allowed.
SYSTEM II Software was unable to
create a histogram queue.
There is insufficient memory to
process the histogram in Multigraph
single histogram display.
File deleted.
Program cannot access Listmode file.
Program cannot access Listmode file.
Program cannot access Listmode file.
You pressed È or Ê during panel
selection. Panel file cannot be found.
Program cannot calculate the available
disk space.
Cytometer failed.
An internal error was detected on
initializing the memory required for
histograms in Multigraph.
An internal error was detected on
initializing the memory used for
histograms in Multigraph.
Panel file is corrupted and cannot be
opened.
TROUBLESHOOTING
ERROR MESSAGES
Action
Try deleting some amorphous regions
before assigning new ones.
Ensure the disk where your
histograms are located is not write
protected. Also, check for a possible
disk full condition.
Exit SYSTEM II Software and restart it.
Enter the Multigraph application. If this
still does not work, call your Coulter
Representative.
Recreate the protocol.
OR
Specify the correct directories on the
Utilities Configuration screen.
OR
From File menu, select Rebuild.
From File menu, select Rebuild.
From File menu, select Rebuild.
From File menu, select Rebuild.
Select panel again.
OR
Specify the correct directories on the
Utilities Configuration screen.
At the Workstation, turn the system
off, and then turn it back on.
Call your Coulter Representative.
Exit SYSTEM II Software and restart it.
Enter the Multigraph application. If this
still does not work, call your Coulter
Representative.
Exit SYSTEM II Software and restart it.
Enter the Multigraph application. If this
still does not work, call your Coulter
Representative.
Re-create the panel or select another
panel.