Beckman Coulter EPICS XL-MCL Troubleshooting Manual page 44

Flow cytometer
Hide thumbs Also See for EPICS XL-MCL:
Table of Contents

Advertisement

REPLACE/ADJUST PROCEDURES
VENTING AIR BUBBLES
3.
Remove the Cytometer center front
panel.
4.
Put the system in the Run or Prime
mode.
5.
Press and hold the vent button until
the bubbles in the vent tubing are
gone.'
6.
Reposition the center front panel. It
snaps into place.
7.
Do the quality control procedure in
the Startup chapter of the
Operator's Guide.
2-6
O.K.
7027031C
7027040C
PN 4237296CA