Table 3.1 Error Messages (Continued)
Message
Invalid Loadlist Panel size changed in
[panel name]
Invalid Loadlist Protocol changed in
[panel name]
Invalid Name. Must be other than
blank characters.
Invalid operator id! Range: (A to Z) and
(0 - 9)
Invalid path selected
Invalid region id
Invalid region index. Call for service.
Invalid user id or password
Invalid value for column '[name]' in
table '[name]'
L
Laser Current Error
Laser Current Okay
Laser Power Error
Laser Power Okay
Laser Regulation Warning
Laser Start Error
Last and only entry in <QC file name>
was erased. File was removed. Erase
corresponding entry in the QCS/QCC
file.
PN 4237296CA
Cause
The Panel size you want to continue
has been modified (protocols have
been added or deleted).
The Panel you want to continue has
been modified. A protocol has been
changed.
You entered a name consisting of
blank spaces.
System cannot accept the entered
characters.
System cannot process the entered
path.
The Region ID entered does not exist
in the current protocol.
Software error.
User has supplied an invalid user ID or
password.
The specified value violates a CHECK
constraint.
The amount of current required by the
laser is outside the expected range.
The amount of current required by the
laser was outside the expected range
but is okay now.
The amount of laser light power is
outside the instrument's operating
range.
The amount of laser light power was
outside the instrument's operating
range but is okay now.
The laser light power is fluctuating.
The laser failed to start within 120
seconds.
After running an _A, _C, or _Q type
protocol, the user erased the last entry
after the protocol finished running.
TROUBLESHOOTING
ERROR MESSAGES
Action
Reload the Panel.
Reload the Panel.
Enter another name.
Enter a valid operator ID.
Enter a valid path.
Specify a region that exists in the
current protocol or create a new
region.
At the Workstation turn the system off,
and then turn it back on.
Enter a valid user ID or password.
Enter the correct value.
Call your Coulter Representative.
Call your Coulter Representative.
Call your Coulter Representative.
Call your Coulter Representative.
Call your Coulter Representative.
Record an entry in the instrument log.
Power the instrument off then on. If
problem persists, call your Coulter
Representative.
This message tells the user that the QC
file had only one entry and since it was
erased, the QC file was deleted. For _A
and _C erase the QCS file (baseline)
and for _Q erase the QCC file.
3
3-15