Beckman Coulter EPICS XL-MCL Troubleshooting Manual page 72

Flow cytometer
Hide thumbs Also See for EPICS XL-MCL:
Table of Contents

Advertisement

TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Error validating your entry
ERROR writing to file [file name]!
Possible full disk
Error writing to log file
F
'Fm__system' not allocated
Fail to close Listmode file
Fail to close Listmode file, Carousel
aborted on tube #[tube number]
Fail to close Listmode file, Current
operation aborted
Failed to read MCL state
Failed to update password
Failure to resume panel: [panel name]
Failure to resume Protocol: [protocol
name]
[Filename] does NOT exist
3-12
Cause
An internal error occurred when
creating a minimum counter.
Software cannot save information to
the named file.
The database engine receiving an I/O
error writing the log file.
Internal error.
Hard disk error. Full disk.
Hard disk error. Full disk.
Hard disk error. Full disk.
MCL status is unknown.
When user attempted to change an old
password in the Maintenance screen
for a new one, the file could not be
found or was damaged.
MCL could not continue with the
current panel.
MCL could not continue with the
current protocol. (The MCL was
running a single protocol.)
The selected file does not exist under
the current directory path.
Action
1. Exit SYSTEM II Software.
2. Call your Coulter Representative.
Specify the correct directories on the
Utilities Configuration screen and
use archive on the Utilities
Configuration screen and free up disk
space.
1. Ensure the disk is not full.
2. Ensure the log file name is valid.
1. Exit SYSTEM II Software.
2. Call your Coulter Representative.
Use archive on the Utilities
Configuration screen and free up disk
space.
Use archive on the Utilities
Configuration screen and free up disk
space.
Use archive on the Utilities
Configuration screen and free up disk
space.
1. Repeat the operation.
2. If the error persists, power off the
Cytometer and restart the SYSTEM
II Software program.
First time use of the Supervisor box in
the Maintenance screen creates the
password file. If this is not the case,
restart the SYSTEM II Software. If
problem persists, get out of the
SYSTEM II program and reboot the
computer.
1. Try to continue with MCL.
2. If the error persists, restart the
SYSTEM II Software program.
1. Try to continue with MCL.
2. If the error persists, restart the
SYSTEM II Software program.
Ensure the directory path is correct.
PN 4237296CA

Advertisement

Table of Contents
loading

This manual is also suitable for:

Epics xl

Table of Contents