Table 3.1 Error Messages (Continued)
Message
VDI display device driver not loaded
properly.
W
Warning: Less than 4 megabytes of
disk space available
Warning: Less than 1 megabyte of disk
space available
Warning: No valid histograms were
noted
Warning: No valid signals were noted
Waste Backpressure Error
Waste Backpressure Okay
Waste Level Error
WASTE LEVEL FULL! Empty tank.
When cyto is READY, press RUN
Waste Level Okay
Waste Level Warning
Worklist file not found
Worklist type panel may not be edited
PN 4237296CA
Cause
No VDI graphics are loaded.
Less than 4 MB is free on the hard
disk.
Less than 1 MB is free on the hard
disk.
You created a protocol with no
histograms.
You created histograms that do not
match the gates you created.
The protocol (in Listmode data
analysis) does not match the signal.
The filter on the waste tank vent line is
probably wet or clogged or
disconnected.
There was too much pressure in the
waste tank but it is okay now.
There is not enough empty volume in
the waste tank for further sample
analysis.
Waste container is full. Waste sensor
failed.
Waste tank for is back to normal from
a waste level warning or error state.
There is enough empty volume in the
waste tank for about 5 minutes of
sample analysis.
Waste sensor failed.
The given Worklist file was not found
at the directory paths as given in the
Utilities Data Management screen.
Tried to select 'Current Worklist' Panel
from the Select menu.
TROUBLESHOOTING
ERROR MESSAGES
Action
Put the software program diskette in
drive A.
At the A:\ prompt, type install.
Use archive on the Utilities
Configuration screen and free up disk
space.
Use archive on the Utilities
Configuration screen and free up disk
space.
Create another protocol.
Create or select another protocol.
OR
Select another Listmode file.
Check waste vent filter for presence of
liquid and proper connection. Call your
Coulter Representative.
None.
Empty the waste container.
Empty the waste container. If the waste
container is not full and this error
appears, call your Coulter
Representative.
None. Informational only.
Empty waste tank.
If the waste tank is not full when this
error message appears, call your
Coulter Representative.
Ensure the directory paths are correct
and repeat the operation.
Do not select Worklist type panel.
3