TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Restarting MCL because: MCL/Panel
panel name mismatch at entry
[carousel entry num]
Restarting MCL because: MCL/ Panel
protocol mismatch at entry [carousel
entry num]
Restarting MCL because: Num of
protocols in [panel file] was changed
Run time SQL error -- [name]
S
Sample Pressure Error
Sample Pressure Okay
Sample pressure warning
Sample Stage Error
Sample Tube Error
3-24
Cause
The MCL could not continue because
the panel name corresponding to the
protocol at the given entry in the
Carousel file is different from the
corresponding entry in the panel file.
The MCL could not continue because
the protocol name at the given entry in
the Carousel file is different from the
corresponding entry in the panel file.
The MCL could not continue because
the number of protocols recorded in
Carousel file corresponding to the
panel at the time it was run is different
from the number of protocols in the
panel as is now.
Watcom SQL has detected a problem.
There may be a leak caused by a bad
sample tube or a bad sample head.
Bad regulator or tubing.
The sample pressure was outside the
system's operating range but is okay
now.
Sample tube not pressurized because
it is bad.
The manual sample stage failed to
move.
On Restart or Continue after Pause, the
system could not sense a sample tube
in the sample stage.
The system sensed a sample tube but
after raising the sample stage it could
no longer sense one.
The system cannot run a cleanse cycle
because there is a sample tube in the
sample stage.
Action
The only choice is to reload the panel
or run the panel from the beginning
using the MCL.
The only choice is to reload the panel
or run the panel from the beginning
using the MCL.
The only choice is to reload the panel
or run the panel from the beginning
using the MCL.
This problem should be reported to
Watcom and/or your Coulter
Representative.
Inspect sample tube and sample head
for damage. Change as required. See
Heading 2.8, Replacing the Sample
Head.
Call your Coulter Representative.
None.
Put sample into another tube.
Call your Coulter Representative.
Reinsert the sample tube.
Remove then reinsert the sample tube.
Remove the sample tube from the
sample stage.
PN 4237296CA