CLEANING PROCEDURES
CLEANING THE SAMPLING SYSTEM
4.
Remove the adaptor and press the
CLEANSE button.
When the cleanse cycle ends, the
CLEANSE button indicator turns
off and the RUN button indicator
flashes. The Cytometer status
message reads Press Run to Initialize.
5.
Record that the vacuum line
cleaning and cleanse cycle were
done on the QC Maintenance
screen. See Appendix B for
instructions.
1-26
7021-056
PN 4237296CA