TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Not enough space for COPY
Not enough space for MOVE
Not enough space to add panel
Not Yet Implemented
O
Only one more prism parameter can be
assigned
Operator ID: <operator ID> is already
assigned
Out of MEMORY printing cytosettings
Out of range
Out of range: 0 - 9,999,999
P
Panel contains deleted protocols or
Error accessing panel file.
Panel has auto setup protocols
Panel has been modified since last
MCL run
[Panel name]: Invalid panel selected
3-20
Cause
There are insufficient data entry lines
on the Worklist to copy the selected
panel.
There are insufficient data entry lines
on the Worklist to move the selected
panel.
There are insufficient entry lines in
Worklist for the selected panel to be
added.
An internal error has occurred.
The system can only accept 4 prism
parameters. 3 are already assigned.
When assigning an operator ID to one
of the operator ID boxes in the
Maintenance screen (located to the
right of the QC Maintenance table) one
of these boxes already contained that
ID.
SYSTEM II Software could not print
the cytosettings due to insufficient
memory.
The system cannot process the
entered value.
The system cannot process the
entered value.
The selected panel contains the name
of protocols that are not present under
current directory path.
Panel contains protocols whose user
name starts with an underscore "_".
These are protocols used for
Autosetup and may not be run in a
Worklist.
The system has detected a modified
panel since the last MCL run.
You selected an invalid panel.
Action
Ensure that the selected panel(s)'s
protocols do not exceed the number of
available data entry lines.
Ensure that the selected panel's
protocols do not exceed the number of
available data entry lines.
Ensure that the selected panel's
protocols do not exceed the number of
available data entry lines. Add the
panel to another area of Worklist or
make room by rearranging the existing
panels in the Worklist.
Exit SYSTEM II Software and call your
Coulter Representative.
Assign one more parameter to prism
or edit the current equation.
Either assign a different operator ID or
change the ID in the box that is in
conflict with ID being assigned.
Exit SYSTEM II Software and call your
Coulter Representative.
Enter a value within the correct range.
Enter a value within specified range.
Edit the Panel to include Protocols in
the current directory or copy the
required protocols into the current
directory.
Select protocols that are NOT
Autosetup. That is, they do not begin
with an underscore "_".
The system restarts the panel from the
beginning.
From File menu, select Rebuild.
PN 4237296CA