REPLACE/ADJUST PROCEDURES
REPLACING THE SHEATH FILTER
CAUTION
fluid to flow in one direction only. Make sure you install the new sheath filter correctly.
3.
Pick up the old sheath filter, and
notice how the three tubes are
connected. Turn the filter and notice
the direction of the arrow on it.
4.
Get the new filter and hold it with
the arrow going in the same
direction as the arrow on the old
filter. In the next step, immediately
install the new filter to avoid spills.
5.
Disconnect and reconnect each tube
to the new filter, one at a time,
beginning with
disconnected by pushing in on the
metal clips on the connectors.
When reconnected, the connectors
snap into place.
6.
Wipe up any spills, and then put the
filter in the drawer.
Check that:
The vent port is above the
r
connector and pointing toward
you.
The tubing is not kinked or
r
twisted.
7.
Record that the sheath fluid filter
was replaced on the QC
Mainenance screen. See Appendix B
for instructions.
2-14
Damage to the instrument can occur if you do not install the sheath filter correctly. It allows
B
. The tubes are
3
2
NEW
SHEATH
FILTER
PUSH
HERE
7027055C
OLD
SHEATH
FILTER
1
3
1
2
VENT
PORT
CONNECTOR
7027107C
PN 4237296CA