Table 3.1 Error Messages (Continued)
Message
[Filename] does not exist in this
directory!
File name is <QC file name>
FILE [filename] REMOVED
File larger than 32768 bytes
File problem
File Write Error [file name]
File Write Error - <QC file name>
H
Help is not available at this time
High channel value must not be
smaller than Low Channel
Host command invalid
I
I/O error [name] -- transaction rolled
back
ID must consist of 12 alphanumeric
characters
Ignoring invalid regions
Illegal Cytometer state
PN 4237296CA
Cause
Error opening the current panel file.
The file being translated to a test file
(used with a spreadsheet program) is
not a QC file.
QC file has been removed.
File could not be exported or sent
through COM port since it exceeded
the maximum allowable size.
Damaged QC file.
Damaged QC file.
The QC file currently being updated
had a write error.
No help screens are available yet.
System cannot accept a numeric
region when the High channel value
entered is smaller than the Low
channel value.
The Cytometer software could not
process the requested command from
the Workstation.
®
Watcom
SQL has detected a problem
with your hard disk.
System cannot accept the characters
entered.
When editing an equation, an invalid
region has been entered.
Cytometer error.
TROUBLESHOOTING
ERROR MESSAGES
Action
1. Exit SYSTEM II Software.
2. Restart the program.
3. If error persists, reboot the
Workstation.
Check the file to be used as a source
file for the translation and make sure it
is a QC file (with QCS or QCC
extension).
Rerun panel to re-create the QC file.
No action.
1. Delete the QC file.
2. Rerun the panel to re-create the QC
file.
1. Delete the QC file.
2. Rerun the panel to re-create the QC
file.
Run the protocol again. Run the panel
if protocol located in a panel. Make
sure the QC path is correct and the QC
file being processed is located in this
path.
None. Informational only.
Enter a value within the range
indicated.
Record an entry into the instrument
log.
Check hard disk using the operating
system or call your Coulter
Representative.
Enter a valid number consisting of
alphanumeric characters.
Verify that the equation entered is the
required equation.
At the Workstation turn the system off,
and then turn it back on.
3
3-13