TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Error in writing file
Error in writing file, Carousel aborted
on tube # [tube number]
Error in writing file, Current operation
aborted
Error initializing protocol
Error installing clock interrupt vector
Error installing Cytometer interrupt
vector
Error installing interrupt [hexadecimal
number] handler
Error: Invalid path selected
Error loading cytometer program file
Error loading file
ERROR LOADING LISTMODE FILE
ERROR LOADING PROTOCOL
3-10
Cause
You specified the wrong path.
Full disk.
System cannot save file to the local
path. Possible full disk.
System cannot save file to the local
path. Possible full disk.
An internal error initializing the
memory for histograms was detected.
SYSTEM II Software cannot properly
initialize itself.
Cytometer interrupt has not been
installed correctly.
SYSTEM II Software cannot properly
initialize itself.
You entered an invalid path.
Software unable to program the
Cytometer.
System cannot load the selected file.
Program cannot access Listmode file.
Program cannot access protocol file.
Action
Specify the correct directories on the
Utilities Configuration screen.
OR
Use archive on the Utilities
Configuration screen and free up disk
space.
At the Utilities Configuration screen,
specify the correct directories.
OR
Use Archive to free up disk space.
At the Utilities Configuration screen,
specify the correct directories.
OR
Use Archive to free up disk space.
1. Exit the SYSTEM II Software.
2. Restart the software.
3. Enter the Multigraph application.
4. If the problem persists, call your
Coulter Representative.
1. Power off the Workstation.
2. Wait a few minutes, then turn the
power on again to the Workstation.
3. If error persists, call your Coulter
Representative.
1. Restart the SYSTEM II Software.
2. If error persists, reset the
computer then restart SYSTEM II
Software program.
1. Power off the Workstation.
2. Wait a few minutes, then turn the
power on again to the Workstation.
3. If error persists, call your Coulter
Representative.
Enter another path.
At the Workstation turn the system off,
and then turn it back on.
1. Repeat the operation.
2. If error persists, then the file is
either missing or is damaged.
From File menu, select Rebuild.
From File menu, select Rebuild.
PN 4237296CA