Beckman Coulter EPICS XL-MCL Troubleshooting Manual page 49

Flow cytometer
Hide thumbs Also See for EPICS XL-MCL:
Table of Contents

Advertisement

WARNING
waste container and its associated tubing might contain residual biological material and must be handled
with care. Clean up spills immediately. Dispose of the contents of the waste container in accordance with
your local regulations and acceptable laboratory procedures
3.
Unscrew the cap and lay it on a
leakproof disposable container,
such as a glove or beaker.
4.
Empty the waste container according
to your laboratory's procedures.
5.
Put about 400 mL of high-quality,
fragrance-free bleach (5% sodium
hypochlorite -available chlorine) in
the waste container to cover the
bottom of the container.
PN 4237296CA
To prevent injury, avoid skin contact with the waste container and its associated tubing. The
REPLACE/ADJUST PROCEDURES
EMPTYING THE WASTE CONTAINER
7027049C
7027051C
2
2-11