TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
E
Equation(s) exceed defined limitations Software failure.
ERROR: 2 parameter histograms not
allowed
Error accessing panel file
Error accessing password
Error allocating extended memory
Error building WorkList panel
ERROR: Channel resolution of
histograms are different
Error creating WorkList LoadList
ERROR! DEFAULT is a reserved name You entered "Default" as a name.
ERROR! <your file name> does not
exist in this directory!
Error forming queue
ERROR: Histogram Log decades are
different
Error in acquiring
Error in entry
Error in equation
ERROR IN LOADING DATA
3-8
Cause
You selected a two-parameter
histogram as control or test.
System unable to add the selected
panel to the Worklist.
The file containing the supervisor
password in the Maintenance screen is
either missing or damaged.
Software cannot allocate extended
memory.
Program could not build Worklist.
You selected control and test
histograms with different resolutions.
Program could not create Worklist.
Software cannot find the named file in
the directory.
System program error.
You selected control and test
histograms with different log types.
Cytometer could not acquire data.
System is unable to accept your input. Enter the correct value.
System unable to compute the
equation entered.
System cannot find the Listmode file.
Either it does not exist, it is corrupted,
or does not have enough memory to
load.
Action
At the Workstation turn the system off,
and then turn it back on.
Select a one-parameter histogram.
1. Ensure the directory path is
correct.
2. Rebuild the panel directory.
3. Repeat the operation.
First time use of the Supervisor box in
the Maintenance screen creates the
password file.
At the Workstation turn the system off,
and then turn it back on.
1. Ensure all fields in the Worklist
screen are correctly filled out.
2. Ensure all entries used have an
assigned protocol.
Select other control and/or test
histograms.
1. Ensure all fields in the Worklist
screen are correctly filled out.
2. Ensure all entries used have an
assigned protocol.
Enter another name.
From File menu, select Rebuild.
Call your Coulter Representative.
Select other control and/or test
histograms.
1. Turn system off at Workstation.
2. Turn the system on again.
1. Verify the validity of the equation.
2. Check that the equation line labels
are lower case letter combinations,
such as: aa, ab, and so forth.
Rebuild the directory or select another
file.
PN 4237296CA