TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Initializing Empty Template
Insufficient memory available
Insufficient memory for dialog box
Insufficient memory for listmode
replay
Insufficient memory to load file
Insufficient memory to load
maintenance
Insufficient space on [drive] : [number
of bytes] bytes free! [number of bytes]
bytes selected!
Internal database error [name]
--transaction rolled back
INTERNAL System ERROR: Invalid
application specified
Invalid count value
Invalid database engine command line It was not possible to start the
Invalid equation string
Invalid Loadlist Attributes changed in
[panel name]
Invalid Loadlist Error accessing panel
[panel name]
Invalid Loadlist Error opening panel
[panel name]
3-14
Cause
Program did not find a template for
this panel.
System needs more memory.
An internal error occurred.
System needs more memory to replay
Listmode data
System needs more memory to load
Listmode data.
The data located in the Maintenance
screen could not be loaded from its file
because there was not enough
memory available.
More space is needed on destination
disk drive to copy or move files.
Watcom SQL detected a problem.
Software error.
You entered an invalid histogram stop
count.
database engine.
You entered an invalid histogram
equation.
The Panel you want to continue has
been modified.
System could not open the specified
panel.
File access error.
Action
If this is a new template, just press
Û to acknowledge the message.
If trying to load an existing template,
try rebuilding the directory.
At the Workstation turn the system off,
and then turn it back on.
Exit SYSTEM II Software program and
call your Coulter Representative.
At the Workstation turn the system off,
and then turn it back on.
On the Utilities Configuration screen,
toggle Color to off.
OR
In the Listmode/Protocol screen,
reduce the histogram resolution.
OR
Add more memory to your computer.
Exit the SYSTEM II program and
reboot the computer.
Remove files from the target drive.
Report this problem to Watcom and/or
your Coulter Representative.
At the Workstation turn the system off,
and then turn it back on.
Enter a valid stop count.
Executable file could not be found.
Check environment variables.
Enter a valid histogram equation.
Reload the Panel.
From File menu, select Rebuild.
From File menu, select Rebuild.
PN 4237296CA