Beckman Coulter EPICS XL-MCL Troubleshooting Manual page 96

Flow cytometer
Hide thumbs Also See for EPICS XL-MCL:
Table of Contents

Advertisement

TROUBLESHOOTING
ERROR MESSAGES
Table 3.1 Error Messages (Continued)
Message
Wrong Date format or out of range
Wrong file
Wrong password
Wrong Time format or out of range
X
X and y channel count exceeded 65535 Count too high when calculating a
X channel count exceeded 65535
Y
Y channel count exceeded 65535
You cannot delete the next use entry
You cannot deselect the next use entry The system cannot deselect the 'Next'
You may not make a deleted or errored
entry as next
You must select a protocol.
Z
Zero length file
3-36
Cause
When entering the date, either an
illegal date was entered or the format
was not as DDMMMYY (for example,
12Apr98).
An error has occurred accessing the
selected configuration.
When trying to activate the Supervisor
ID box on the QC Maintenance screen,
the user has entered the wrong
supervisor password.
The system cannot process the Time
value entered.
projection.
Count too high when calculating a
projection.
Count too high when calculating a
projection.
The system cannot delete the 'Next'
used entry from a queue.
used entry.
The system can assign 'Next' to a
deleted protocol.
The system cannot assign a deleted or
corrupted protocol as 'Next'.
You must first select a protocol to edit
this column.
File is empty.
Action
Enter correct format and make sure
the day does not exceed max month
day.
Rebuild configuration directory and
retry selecting the configuration file. If
unsuccessful, call your Coulter
Representative.
Enter the correct password. If
supervisor is unable to reproduce the
correct password then the Master
Password must be used. This
password should be in possession of
an authorized person.
Enter a valid time.
Statistics are invalid
Statistics are invalid.
Statistics are invalid.
To delete that entry, reassign the 'Next'
to another entry and then delete.
To deselect the 'Next' entry, reassign
'Next' first.
Do not assign 'Next' to a deleted
protocol.
Reassign 'Next'.
Select a protocol in the
protocol/equation column.
Ensure file for export contains data.
PN 4237296CA

Advertisement

Table of Contents
loading

This manual is also suitable for:

Epics xl

Table of Contents