REPLACE/ADJUST PROCEDURES
REPLACING THE MANUAL SAMPLE HEAD
2.
At the sample stage, remove the two
Allen screws securing the test tube
guide around the sample head.
3.
Remove the test tube guide and set
it aside.
4.
Pull down on the sample head while
turning it to the left and slide it off
the sample probe.
5.
Transfer the tubing from the old
head to the new head one at a time.
2-16
7027118C
PN 4237296CA