12.4 Calibrating the Touch Screen

The iNEX has the touch screen calibrated at the factory, however over
time the touch sensor may change and require recalibration.
1. Start the iNEX into Launcher mode
(see 12.1 Starting Up in Launcher
Mode).
2. Press CALIBRATE.
3. Press each of the on-screen targets as
they are shown. When complete press
ACCEPT.

12.5 System Recovery Options

In the event that the iNEX becomes unresponsive or data on the system
becomes unusable there are a number of steps that can be taken to
recover the iNEX.
A forced shutdown will make unresponsive systems turn off. In most
cases when the iNEX is then turned on again the problem will be resolved.
If a forced shutdown is often required the system may need servicing.
If data on the system becomes unusable then a corruption may have
occurred. There are a number of options to fix data corruptions and each
should be tried in the following order:
1. Check Disk
2. Repair Database
3. Soft Reset
4. Hard Reset.
The Soft Reset and Hard Reset options should be used as a last resort as
they will delete some or all data from the iNEX.
iNEX User Manual
Page 93
Rev 2