Equipment Required - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

3
Testing Performance

Equipment Required

Table 9

Equipment Required

Equipment
Pulse Generator
150 ps Transition Time Converter
(Qty 3)
Oscilloscope
SMA Coax Cable
(Qty 2)
Flying Lead Probe Set with 5 ground
leads
(Qty 2)
Male BNC to Female SMA adapters
(Qty 2)
SMA/Flying Lead test connectors, (f)
SMA to (f) SMA to Flying Lead Probe
(Qty 3)
34
The following equipment is required for the performance test
procedure.
Critical Specification
≥ 260 MHz, two channels, differential
outputs, 150-180 ps rise/fall time (if
faster, use transition time converters)
Required if pulse generator's rise time
is less than 150 ps. (Pulse generator
conditions: Voffset=1V, ΔV=500 mV.)
Required for 8133A opt. 003.
bandwidth ≥ 1.5 GHz,
sampling rate ≥ 8 GSa/s
>18 GHz bandwidth
no substitute
no substitute
Recommended Model/Part
Agilent or HP 8133A option 003
Agilent or HP 15435A
Agilent or HP 54845A/B or similar
Agilent or HP 8120-4948
Agilent or HP E5383A
Cambridge Products CP-AD507
(see www.cambridgeproducts.com)
See
"To Assemble the SMA/Flying
Lead Test Connectors"
16800 Series Portable Logic Analyzers Service Guide
on page 28