Connect the Test Equipment

Connect the Logic Analyzer Pod to the 8133A Pulse Generator

16800 Series Portable Logic Analyzers Service Guide
Connect a Transition Time Converter (if required—see
1
Table
7, "Recommended Test Equipment," on page 19) to
the 8133A pulse generator's Channel 2 OUTPUT, Channel
2 OUTPUT, and Channel 1 OUTPUT.
2
Connect the SMA/Flying Lead test connector (see
Assemble the SMA/Flying Lead Test Connectors"
page 28) with the 50 ohm terminator to the Transition
Time Converter at the 8133A pulse generator Channel 1
OUTPUT. (If Transition Time Converters are not required,
connect the SMA/Flying Lead test connector directly to
the pulse generator output.)
Connect the two SMA/Flying Lead test connectors without
3
50 ohm terminators to the Transition Time Converters at
the 8133A pulse generator Channel 2 OUTPUT and
Channel 2 OUTPUT. (If Transition Time Converters are
not required, connect the SMA/Flying Lead test
connectors directly to the pulse generator outputs.)
4
Connect an E5383A Flying Lead Probe Set to Pod 1 of the
logic analyzer.
Connect the E5383A Flying Lead Probe Set's CLK lead to
5
the pin strip of the SMA/Flying Lead connector at the
8133A pulse generator's Channel 1 OUTPUT.
ground lead
3
Testing Performance
"To
on
39