To remove and replace the front panel assembly

16800 Series Portable Logic Analyzers Service Guide
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 120.
"To remove and replace the cover"
Disconnect the following cables:
2
USB cables from the motherboard.
Display cable from the display board (use care when
disconnecting to avoid damage).
Optional touch screen cable from the touch screen
controller board.
Keyboard cable from the keypad board.
Front panel cable from the backlight inverter board.
Using a Torx T10 screwdriver, remove 6 screws that
3
secure the front panel to the chassis.
4
Remove the front panel assembly.
5
Reverse this procedure to install the front panel assembly.
T-10
screws (3)
USB
cables
Display
cable
Replacing Assemblies
on page 122.
Keyboard
cable
Front panel
cable
Touch screen
cable
6
on
T-10
screws (3)