To remove and replace the front panel assembly
16800 Series Portable Logic Analyzers Service Guide
1
Perform previous procedures:
•
"To prepare the instrument for disassembly"
page 120.
•
"To remove and replace the cover"
Disconnect the following cables:
2
•
USB cables from the motherboard.
•
Display cable from the display board (use care when
disconnecting to avoid damage).
•
Optional touch screen cable from the touch screen
controller board.
•
Keyboard cable from the keypad board.
•
Front panel cable from the backlight inverter board.
Using a Torx T10 screwdriver, remove 6 screws that
3
secure the front panel to the chassis.
4
Remove the front panel assembly.
5
Reverse this procedure to install the front panel assembly.
T-10
screws (3)
USB
cables
Display
cable
Replacing Assemblies
on page 122.
Keyboard
cable
Front panel
cable
Touch screen
cable
6
on
T-10
screws (3)
129