To remove and replace the pattern generator probe cable

16800 Series Portable Logic Analyzers Service Guide
Use a Torx T10 screwdriver to remove the screws that
1
hold the cable clamp to the back panel of the card.
Remove the faulty probe cable from the connector and
2
install the replacement cable.
Install the label on the new probe.
3
If you order a new probe cable, you also need to order
new labels. Probe cables shipped with the card are
already labeled. For part numbers and ordering
information, see
"Pattern Generator Card Replaceable
Parts"
on page 167.
Install the 4 screws that secure the cable clamp to the
4
back panel of the card.
5
Reverse this procedure when replacing the cards.
The pattern generator card goes in the lower position.
Seat the lower card first.
a
Press firmly to ensure the cards are fully seated in the
b
connector on the module interface board.
Replacing Assemblies
T-10
screws
(4)
6
143