To Remove And Replace The Cover - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

6
Replacing Assemblies

To remove and replace the cover

T-20 screws (2)
CAU TI O N
122
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 120.
Using a Torx T20 screwdriver, remove the handle
2
assembly.
3
Using a Torx T10 screwdriver, remove 6 screws that
secure the cover to the chassis.
Slide the cover back to remove.
4
Reverse this procedure to replace the cover.
5
When reinstalling the handle assembly, ensure that the screws are
torqued to 2.372 Newton meters (21 inch pounds) so that they do
not work themselves loose.
T-10 screws (6)
16800 Series Portable Logic Analyzers Service Guide