Secondary Power Supply - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

6
Replacing Assemblies
Release latches on
600 watt power supply
cable
Release tab on
600 watt power
supply cable

Secondary power supply

124
MIB
Using Torx T20 screwdriver, remove the 4 screws that
5
secure the 600 watt power supply to the chassis.
6
Slide the power supply toward the front panel and lift it
out of the frame while holding the line filter cable for the
lower power supply out of the way.
7
Reverse this procedure to replace the 600 watt power
supply.
Depending on the model you ordered, your secondary power
supply may be 175 watts or 15 watts and may look different
than the one shown here.
1
Perform the procedures:
"To prepare the instrument for disassembly"
page 120.
"To remove and replace the cover"
"To remove and replace the front panel assembly"
page 129.
2
Using a Torx T10 screwdriver remove the 2 side screws
securing the tray assembly to the frame.
Slide the tray out far enough to disconnect the following
3
cables:
Release latches on line
filter cable
T-15
screws (4)
16800 Series Portable Logic Analyzers Service Guide
600 watt
power supply
Line filter cable for
lower power supply
on
on page 122.
on