To remove and replace a measurement circuit board

16800 Series Portable Logic Analyzers Service Guide
Remove the 4 screws attaching the ground spring and
1
back panel to the circuit board, then remove the back
panel and the ground spring.
Replace the faulty circuit board with a new circuit board.
2
On the new board, ensure the 2x15 (30- pin) ribbon cable
is connected between J15 and J12.
Position the ground spring and back panel on the back
3
edge of the replacement circuit board. Install 4 screws to
connect the back panel and ground spring to the circuit
board.
4
Install the logic analyzer cables using the procedure
remove and replace measurement cables"
Install the logic analyzer card using the procedure
5
remove and replace a measurement card"
Replacing Assemblies
T-10
Screws
(4)
Back
panel
Ground
spring
Circuit
board
"To
on page 139.
"To
on page 138.
141