Set up the test equipment

Table 14 Pulse Generator Setup for Testing Cables
Timebase
Pulse Channel 2
Mode: Pulse ÷ 1
Mode: Int
Freq: 40 MHz
Delay: (not available in pulse mode)
Width: 4 ns
Ampl: 3 V
Offs: 1.5 V
Output: Enable (LED off)
Comp: Normal (LED off)
Limit: Off (LED off)
Output: Enable (LED off)

Connect the test equipment

16800 Series Portable Logic Analyzers Service Guide
If you have not already done so, do the procedure
1
run the self- tests"
Set up the pulse generator.
2
Set up the pulse generator according to the following
a
table:
Trigger
Disable (LED on)
1
Using two 2 x 9 test connectors, connect the logic
analyzer to the pulse generator channel outputs.
Connect the even- numbered channels to the pulse
a
generator Channel 2 OUTPUT.
b
Connect the odd- numbered channels the pulse
generator Channel 2 OUTPUT.
Connect Clk1 to the pulse generator Channel 2 OUTPUT.
2
Enable the pulse generator Channel 1 and Channel 2
3
outputs (LEDs off).
on page 88.
Pulse Channel 1
Doesn't matter, not used in this test.
5
Troubleshooting
"To
105