Recommended Test Equipment - Agilent Technologies 16800 Series Service Manual

Portable logic analyzers
Hide thumbs Also See for 16800 Series:
Table of Contents

Advertisement

Recommended Test Equipment

Table 7
Recommended Test Equipment
Equipment
Single-ended Flying Lead Probe Set
(Qty 2)
Ground Leads (Qty 5)
Pulse Generator
150 ps Transition Time Converter
(Qty 4)
Oscilloscope
SMA/Flying Lead Test connectors
2 x 9 Test connectors
SMA Coax Cable (Qty 2)
†P = Performance Tests, T = Troubleshooting
16800 Series Portable Logic Analyzers Service Guide
Critical Specifications
no substitute
no substitute
260 MHz,1 ns pulse width, two channels,
≤ 150 ps rise time
Required if pulse generator's rise time is less
than 150 ps (Voffset=1V, ΔV=600 mV).
Required for 8133A opt. 003
≥ 1.5 GHz bandwidth,
≥ 8 GSa/s sampling rate
no substitute
no substitute
≥ 18 GHz bandwidth
General Information
Recommended Agilent
Model/Part
E5383A
pkg of 5
(Included with E5383A
Probe Set)
8133A Option 003
Agilent or HP 15435A
54845A or 54845B
See
"To Assemble the
SMA/Flying Lead Test
Connectors"
on page 28
See
"To assemble the 2 x 9
test connectors"
on
page 102
8120-4948
1
Use†
P, T
T
P, T
P
P
P
P
P
19

Advertisement

Table of Contents
loading